| Capacity
Available |
90 Litres to 8000 Litres |
| Temperature
Range |
Ambient to 80 C
from -20C to 80 C |
| Accuracy |
+/- 1C or 2C |
| Humidity
range |
45% RH to 98% RH over a temp. tange of 20C
to 85C |
| Reference
Standard |
IS 9000, JSS 55555, JSS 50101,
IEC 68, DIN, DEF, ASTM and Departmental standards
like QM 333, DOE, C-DOT and as per customers specification for cold tests. |
| SALIENT
FEATURES |
| Interior
Material |
18 swg 304 S.S. |
| Outer
Material |
18 swg CRCA with electrostatic power coating Paint. |
| Insulation |
4" / 6" thick class 'A' fibre glass wool or polyurethane
foam. |
|
Air
Circulation |
Suitable capacity of fan motor with suitable fan blade to maintain the
air velocity between 0.5 to 2 mtrs./sec |
| Door |
Double walled, full front opening door with suitable locking
arrangement. |
|
Gasket |
Specially designed ,double lined silicon gasket will be
provided |
|
Controller |
Solid state digital / PID/ Microprocessor temp.
and humidity (direct % RH or dry and wet
bulb) indicator cum controller with PT -100 sensing
|
| Refrigeration |
Suitable capacity of single stage/ 2 stage cascade with suitable
refrigerant or CFC free gases. The chambers will be completely Frost Free chambers |
| Heating
System |
Suitable capacity of air heaters inconel /S.S. tubes will be
provided |
| Humidity
System |
Suitable capacity of steam injection/vapour injection with automatic
water level controller will be provided
|
|
Safety |
Low and high voltage protection and low and high
temp. protection will
be provided |
|
Utilities |
Port holes, trays, terminal panels, inspection lamp viewing window, any
other utilities as per customers' requirements will be provided.
|
| Power
Supply |
230 v, Single phase,50 Hz or 415 v, 3 phase, with neutral 50 Hz |